Connect the ground from the external sensor supply to the GND connection on the LMA-
3
Connect the +5V from the external sensor supply to the ‘+’ input of the channel to be
calibrated
Press the Channel button until the channel number to be calibrated is displayed
Press and hold the Calibrate button while the channel number is still displayed
The LMA-3 will either display a lower case ‘c’ to indicate the Calibration has occurred for the
selected channel or an upper case ‘E’ to indicate that the calibration could not occur. Normally
an E indicates that the voltage being input is .1V or more off from 5V.
10.1.3 Restoring a channel to factory calibration
In order to restore an external input back to the factory default calibration, the following steps
must be taken.
Connect a jumper wire from the ‘+’ input of the channel to be restored to the GND
connection on the LMA-3